Trouble-Shootings
Example of troubleshooting for ELISA kit
Troubleshooting Guide
| Problem | Possible causes | Solution |
|
Poor standard curve |
·       Inaccurate pipetting
·       Insufficient vortexing ·       OD450 too high for the high standard point ·       Air bubbles in wells. ·       Standard defect or not fully recovered |
·       Check pipette calibration and ensure equal dispensing.
·       Vortex 30 sec for each of standard dilution steps and vortex again (10 sec) before pipetting to the 96-well plate. ·       Reduce substrate incubation time ·       Remove air bubbles in wells by pipette tip. ·       Change a standard vial or spin down the vial before reconstitution |
|
Low signal |
·       Improper preparation of reagents and storage
·       Too brief incubation times ·       Inadequate reagent volume or improper dilution ·       Standard defect and sample overdiluted |
·       Spin down vials before opening. Reconstitute the content thoroughly. Proper storage of plate and strip and detection antibody after first usage.
·       Microplate shaker may improve signals. ·       Insufficient HRP Conjugate. Ensure sufficient incubation time and increase sample incubation to 2 h. ·       Change a Standard vial. Undilute sample or less dilution |
| Overflow in the standards | ·       Substrate incubation too long
·       Air bubbles in wells |
·       Observe the color development every 1-2 mins and reduce substrate incubation time.
·       Stop the reaction by adding 50 µl of Stop Solution when it turns to dark blue in the highest concentration of standard wells. ·       Remove air bubbles in wells |
|
Large CV |
·       Inaccurate pipetting and mixing
·       Improper standard/sample dilutions. ·       Air bubbles in wells. ·       Microplate reader out of calibration ·       It did not turn yellow after adding Stop Solution |
·       Check pipettes and ensure the pipette is calibrated properly.
·       Ensure accurate pipetting and thorough mixing. ·       Use reverse, instead of forward pipetting. ·       Use the correct dilution buffers ·       Remove air bubbles in wells by pipette tip. ·       Calibrate the microplate reader properly ·       If it did not turn yellow after adding Stop Solution, mix with pipette tip till it turns yellow prior to measurement. |
|
High background |
·       Reagent reservoir issue
·       Plate is insufficiently washed and air bubbles in wells. ·       Contaminated Assay Buffer ·       Pipet tip contaminated |
·       Use a new reagent reservoir for Substrate Solution.
·       Increase wash to 4 times before adding substrate and ensure plate washer functions normally. Remove air bubbles in wells by pipette tip. Use squirt bottle for washing. ·       Make fresh Assay Buffer and wash thoroughly. ·       Use new pipette tips for blank wells. |
|
No signal detected |
·       The procedure was misconducted.
·       Failures of spin down the contents in Detection Antibody and Standards. ·       Failure of HRP or Substrate Samples overdiluted |
·       Ensure the step-by-step protocol. Spin vials of Detection antibody and Standard to completely recover the content.
·       Ensure HRP volume. Mix 100 µl of Substrate with 0.5 µl HRP and dark blue color should develop in 5 min. ·       Try a new standard vial and use positive control. ·       Try not dilute samples |
|
Low sensitivity |
·       Improper dilutions of standards
·       Improper storage of the ELISA kit |
·       Ensure accurate and thorough dilutions of standards at each step.
·       Store detection antibody at -20°C after reconstitution and others at 4°C. Keep substrate solution protected from light. |